This precision analyzer has a range from 0 to 100%. It is typically used to determine the O2concentration in respired gases or gas mixtures from collection bags, or
other sources.
The O2 concentration is displayed on a digital meter. With the end-tidal (EtO2) option the lowest value during a breathing cycle may be displayed. Corresponding analog voltages are available on rear panel BNC connectors.
A gas pump is built in to draw the gas through the analyzer.
A flow meter indicates the gas sample flow. Gas flow is internally adjustable. Adjustment may affect response time. A gas return port on the rear panel is standard.
Precision gas mixtures are available for calibrating the analyzer.
Meter
Display: |
|
Repeatability: |
|
Linearity: |
|
Zero Stability: |
(At Constant Temperature
& Constant Flow) +/- 0.002% O2/hour
|
Temperature Coefficient: |
(Within a range of +/-10
degree C from calibration temperature)
Zero +/- 0.03+/- O2/ degree C
Typical Span +/- 0.1%
FS/degree C Typical
|
Response Time T90,
N2 to 100% O2: |
2.0 sec at 400 ml/min flow
|
Type: |
|
Digital Panel Meter
Readout: |
|
BNC Signal Out X1: |
0 - 5 VDC = 0 to 100% O2 0 - 10 VDC on request, no charge
|
BNC Signal Out X4: |
|
Output Noise: |
Less than 2 mv peak to peak
|
Operating Voltage: |
|
Size: |
5-3/4" (146 mm) Height
x 8-5/8" (229 mm) Width x 15" (380 mm) Depth
|
Sample Gas Inlet: |
Rear Panel (Front panel on request, no charge)
|
Drying Column: |
Nafion semi-permeable membrane
brings sampled gas to ambient humidity. No chemical dryer required.
|
Fast Response Option
Model 17518C-FR uses an optional fuel cell sensor with a response time of about 150 ms. Some of above specifications do not apply.
1 year warranty details in FAQ