This precision analyzer has
a range from 0 to 10%. It is typically used to determine the CO2
concentration in respired gases or gas mixtures from collection bags, or
other sources. May be adjusted to 0 to 15% CO2 on request.
The CO2
concentration is displayed on a digital meter and a corresponding analog
voltage is available on a rear panel BNC connector.
A gas pump is built in to
draw the gas through the analyzer. Gas flow is internally adjustable. Adjustment may affect response time. A gas return port may be installed on request at no charge.
Precision gas mixtures are
available for calibrating the analyzer.
Photo also shows companion
O2 analyzer.
Specifications
Not intended for clinical
use, for teaching or laboratory use only
Type: |
Infrared |
Range: |
0 - 10% |
Sensitivity: |
+/- 0.01%
CO2 |
Resolution: |
0.01%
CO2 |
Zero Drift |
Less than 1.5 torr in 1 hour |
Output: |
0 - 5 Volt = 0 to 10% CO2 |
Meter Display: |
3-1/2
digit LED |
Flow
Rate (nominal): |
200 ml/minute |
Response Time: |
120 milli-seconds typical |
Operating Temperature
Range: |
5o
to 40o C |
External Power Supply: |
100 -
250 VAC, 50/60 Hz (Included), 0.5 Amp at 115VAC |
Warm-up: |
30 minutes
to 0.1% CO2 |
Size: |
2.75" (70 mm) Height x 6.3" (155 mm) Width x 10" (255 mm) Depth |
Drying Column: |
Nafion semi-permeable membrane brings sampled gas to ambient humidity.
No chemical dryer required.
|
Gas Sample Inlet: |
Front
Panel |
See also: Last Updated: 1 Jun 2022